  • Laser Type: Semi-conductor Diode Laser
  • Laser Safety Classification: Class IV (21CFR1040.10) or 4 (EN60825-1)
  • Wavelength: 810nm or 980nm
  • Laser Modes: Continuous
  • Treatment Modes: Continuous, Single or Repeat Pulse
  • Maximum Average Power: 15, 20 or 25 Watts, selectable from 1 Watt to max.
  • Pulse On Time (Exposure Duration): Selectable from 0.05 to 99 seconds
  • Pulse Off Time (Repeat Exposure Interval): Selectable from 0.10 to 99 seconds
  • Fiber Port (Proximal Laser Aperture): SMA-905 connector compatible port
  • Fiber Compatibility: Silica fiber core diameters greater than or equal to 365 microns and Fiber numerical aperture greater than or equal to 0.29 NA
  • Cumulative Energy: Total amount of energy delivered displayed on keypad
  • Cumulative Time: Total amount of lasing time displayed on LCD display
  • Aiming Beam: 635-650 nm, adjustable intensity, either continuous or pulsed
  • Input voltage/Current Rating: 100-240 VAC, 540 VA max., 50/60 Hz
  • Safety Compliance: EN60601-1, UL60601-1, CAN/CSA C22.2 No. 601-1, ETL Listed, EN60601-2-22, EN60825-1, YZB
  • Dimensions: 16″ (40.6 cm) Depth x 9″ (22.8 cm) Width x 14″ (35.6 cm) Height
  • Weight: 24.5 lbs. (11.2 kg)

Standard with each LaserPro Diode are the following items:

  • Fast turn on with no warm up
  • Easy-to-operate user interface speeds learning
  • Small foot print, lightweight, transportable unit.  Optional cart available
  • Foot switch (standard) or finger touch (available with certain fibers) laser control
  • Laser safety signs and Laser safety glasses for physician, patient and nurse included